PCE Instruments PCE-BDM 20 Digital Multimeter for Stationary Workplace

  • Part #:PCE-BDM 20
  • Condition:Brand New
0
Reviews Write a Review Questions & Answers
PCE Instruments PCE-BDM 20 Digital Multimeter for Stationary Workplace
PCE Instruments PCE-BDM 20 Digital Multimeter for Stationary Workplace
PCE Instruments PCE-BDM 20 Digital Multimeter for Stationary Workplace
PCE Instruments PCE-BDM 20 Digital Multimeter for Stationary Workplace
PCE Instruments PCE-BDM 20 Digital Multimeter for Stationary Workplace
  • List Price: $220.00
  • $217.80 / Each
    You Save: 1.0%
  • Free Shipping logoFree Shipping
  • List Price: $220.00
  • $217.80 / Each
    You Save: 1.0%
  • Free Shipping logoFree Shipping
Typically In Stock
Check Availability
Quantity:
Description
See more product information Report an issue with this product
⚠️ Attention CA Residents: Prop 65 Warning
Technical details
AC voltage 
Measuring rangeResolution / Accuracy
60 mV0.001 mV / ± (0.6% 60 digits) at 45 Hz ... 1 kHz
600 mV0.01 mV / ± (0.3% 30 digits) at 45 Hz ... 1 kHz
6V0.0001V / ± (0.3% 30 digits) at 45 Hz ... 1 kHz
60V0.001V / ± (0.3% 30 digits) at 45 Hz ... 1 kHz
600V0.01V / ± (0.4% 30 digits) at 45 Hz ... 1 kHz
1000V0.1V / ± (0.6% 30 digits) at 45 Hz ... 1 kHz
Input impedance:> 10MΩ
Surge protection: 1000V
Display: True RMS in the measuring range of 10 ... 100% of the respective measuring range
DC voltage 
Measuring rangeResolution / Accuracy
60 mV0.001 mV / ± (0.025% + 20 digits)
600 mV0.01 mV / ± (0.025% + 5 digits)
6V0.0001V / ± (0.025% + 5 digits)
60V0.001V / ± (0.025% + 5 digits)
600V0.01V / ± (0.003% + 5 digits)
1000V0.1V / ± (0.003% + 5 digits)
Input impedance:> 10MΩ
Surge protection: 1000V
AC voltage + DC voltage
Measuring rangeResolution / Accuracy
60 mV0.001 mV / ± (1% + 80 digits) at 50 Hz ... 1 kHz
600 mV0.01 mV / ± (1% + 80 digits) at 50 Hz ... 1 kHz
6V0.0001V / ± (1% + 80 digits) at 50 Hz ... 1 kHz
60V0.001V / ± (1% + 80 digits) at 50 Hz ... 1 kHz
600V0.01V / ± (1% + 80 digits) at 50 Hz ... 1 kHz
1000V0.1V / ± (1.2% + 80 digits) at 50 Hz ... 1 kHz
Input impedance:> 10MΩ
Surge protection: 1000V
Display: True RMS in the measuring range of 10 ... 100% of the respective measuring range
Alternating current
Measuring rangeResolution / Accuracy
600 µA0.01 µA / ± (0.6% + 40 digits) at 45 Hz ... 1 kHz
6000 µA0.1 µA / ± (0.6% + 20 digits) at 45 Hz ... 1 kHz
60-mA0.001-mA / ± (0.6% + 40 digits) at 45 Hz ... 1 kHz
600-mA0.01-mA / ± (0.6% + 20 digits) at 45 Hz ... 1 kHz
10 A0.001 A / ± (1% + 20 digits) at 45 Hz ... 1 kHz
Display: True RMS in the measuring range of 10 ... 100% of the respective measuring range
Overload protection: µA and mA Measurement: Fuse F 0.6 A 1000V Ø6 x 32 mm
10 A measurement: F 11 A 1000V Ø 10 x 38 mm
When measuring almost 20 A, the measurement time should not be longer than 30 seconds.
After the measurement, the measuring device must rest for approx. 10 minutes.
Direct current
Measuring rangeResolution / Accuracy
600 µA0.01 µA / ± (0.8% + 20 digits)
6000 µA0.1 µA / ± (0.8% + 10 digits)
60-mA0.001-mA / ± (0.8% + 20 digits)
600-mA0.01-mA / ± (0.15% + 10 digits)
10 A0.001 A / ± (0.5% + 10 digits)
Overload protection: µA and mA Measurement: Fuse F 0.6 A 1000V Ø6 x 32 mm
10 A measurement: F 11 A 1000V Ø10 x 38 mm
A measurement of almost 20 A should not take longer than 30 seconds.
After the measurement, the measuring device must rest for approx. 10 minutes.
AC + DC
Measuring rangeResolution / Accuracy
600 µA0.01 µA / ± (0.8% + 40 digits) at 50 Hz ... 1 kHz
6000 µA0.1 µA / ± (0.8% + 20 digits) at 50 Hz ... 1 kHz
60-mA0.001-mA / ± (0.8% + 40 digits) at 50 Hz ... 1 kHz
600-mA0.01-mA / ± (0.8% + 20 digits) at 50 Hz ... 1 kHz
10 A0.001 A / ± (1.2% + 20 digits) at 50 Hz ... 1 kHz
Display: True RMS in the measuring range of 10 ... 100% of the respective measuring range
Overload protection: µA and mA Measurement: Fuse F 0.6 A 1000V Ø6 x 32 mm
10 A measurement: F 11 A 1000V Ø10 x 38 mm
A measurement of almost 20 A should not take longer than 30 seconds.
After the measurement, the measuring device must rest for approx. 10 minutes.
Resistance
Measuring rangeResolution / Accuracy
600 Ω0.01Ω / In REL mode: ± (0.05% + 10 digits)
6 kΩ0.0001 kΩ / ± (0.05% + 2 digits)
60 kΩ0.001 kΩ / ± (0.05% + 2 digits)
600 kΩ0.01 kΩ / ± (0.05% + 2 digits)
6 MΩ0.0001 MΩ / ± (0.3% + 10 digits)
60 MΩ0.001 MΩ / ± (2% +10 digits)
Surge protection: 1000 V
With a measuring range of 60 MΩ the ambient humidity should be < 50%.
Conductivity
Measuring rangeResolution / Accuracy
60 nS0.01 nS / ± (2% + 10 digits)
Surge protection: 1000V
The ambient humidity for the measuring range should be < 50%.
Capacity measurement
Measuring rangeResolution / Accuracy
6 nF0.001 nF / ± (3% + 10 digits)
60 nF0.01 nF / ± (2.5% + 5 digits)
600 nF0.1 nF / ± (2% + 5 digits)
6 μF0.001 μF / ± (2% + 5 digits)
60 μF0.01 μF / ± (2% + 5 digits)
600 μF0.1 μF / ± (2% + 5 digits)
6 mF1 μF / ± (5% + 5 digits)
60 mF10 μF / for reference only
Surge protection: 1000 V
Temperature
Measuring rangeResolution / Accuracy
-40 ... 40°C1°C / ± (2.0% + 30 digits)
40 ... 400°C1°C / ± (1.0% + 30 digits)
100 ... 1000°C1°C / ± 2.5%
-40 ... 104°F1.8°F / ± (2.5% + 50 digits)
104 ... 752°F1.8°F / ± (1.5% + 50 digits)
752 ... 1832°F1.8°F / ± 2.5%
Surge protection: 1000V
Two-channel temperature measurement
Temperature sensor: Type K only applies to the measurement of temperatures below 230°C / 446°F
Frequency measurement
Measuring rangeResolution / Accuracy
60 Hz0.001 Hz / ± (0.02% + 8 digits)
600 Hz0.01 Hz / ± (0.01% + 5 digits)
6 kHz0.0001 kHz / ± (0.01% + 5 digits)
60 kHz0.00 1kHz / ± (0.01% + 5 digits)
600 kHz0.01 kHz / ± (0.01% + 5 digits)
6 MHz0.0001 MHz / ± (0.01% + 5 digits)
60 MHz0.00 1MHz / ± (0.01% + 5 digits)
Surge protection: 1000V
Input amplitude: 10 Hz ... 30 MHz: 600 mV <a <30 Vrm,> 30 MHz: not specified
Duty cycle
Measuring rangeResolution / Accuracy
10 ... 90% at0.01% / ± (1.2% + 30 digits)
(10Hz ... 2kHz) 
Surge protection: 1000V
Rise time: <1 µs, the signal is centered on the trigger level
Pulse width
Measuring rangeResolution / Accuracy
250 mS0.001 ... 0.1 mS / ± (1.2% + 30 digits)
Surge protection: 1000V
Rise time: < 1 µs, the signal is centered on the trigger level
10 Hz ... 200 kHz: pulse width > 2 µs
The pulse range is limited by the frequency of the signal
Continuity test 
ResolutionFunctional description
0.01 ΩShort circuit alarm: from < 10 Ω, alarm switches off from > 50 Ω
Surge protection: 1000V
Diode test 
0.0001VThe open circuit voltage is approx. 3V and the measurable voltage drop
 of the PN transition is < 3V A continuous beep
 indicates the short circuit of the PN junction. The typical tension
 of the silicon PN junction is between 0.5 ... 0.8V.
Surge protection: 1000V
Further specifications on accuracy can be found in the operating instructions
  
Further specification 
InterfaceUSB interface
Supply voltage100 ... 240V adjustable
Environmental conditions  23°C / 73.4°F ± 5°C / 41°F, < 75% RH
Dimensions310 x 240 x 120 mm / 12.2 x 9.4 x 4.7 in
WeightApprox. 3713 g / 8.2 lbs