Gas Detectors & Analyzers MRU Instruments 690126 AMPRO 1000 Combustion Analyzer O2, CO, NOx - HVAC - ABS Transport Case & Printer

  • Part #:690126
  • Model #:AMPRO 1000
  • Condition:Brand New
51 Review Write a Review
Gas Detectors & Analyzers MRU Instruments 690126 AMPRO 1000 Combustion Analyzer O2, CO, NOx - HVAC - ABS Transport Case & Printer
Gas Detectors & Analyzers MRU Instruments 690126 AMPRO 1000 Combustion Analyzer O2, CO, NOx - HVAC - ABS Transport Case & Printer
Gas Detectors & Analyzers MRU Instruments 690126 AMPRO 1000 Combustion Analyzer O2, CO, NOx - HVAC - ABS Transport Case & Printer
Gas Detectors & Analyzers MRU Instruments 690126 AMPRO 1000 Combustion Analyzer O2, CO, NOx - HVAC - ABS Transport Case & Printer
  • List Price: $1,699.00
    • $1,651.00 / Kit
      You Save: 2.83%
  • Free Shipping logoFree Shipping
See product details
Typically In Stock
Check Availability
Quantity:
Technical details
Operation temperature41 °F … 113 °F
Storage temperature-4°F … 122°F
Ambient conditions95% Rel. Humidity, non-condensing
Data storage1,000 measurements, 1,000 sites
InterfacesMini-USB, Micro-SD, IRDA, Bluetooth (data transfer to Smartphone, Tablet or PC)
Internal Battery PackLithium-Ion battery, 2250 mAh
Grid power supply100 - 240 VCA / 50 … 60 Hz 500mA
Protection classIP30
WeightApproximately 1.1 lbs. (with 2 sensors)
Dimensions(W x H x D) 3.23" x 6.65" x 1.73"
ApprovalsTUV-certified according to 1. BlmSchV and EN 50379-1 and 2 (Version B)

or EN 50379-3 (Version E), ByRgG 309

Calculated valuesRange
Carbon dioxide0 … CO2 max.
Heat losses qA0 … 99.9 %
Efficiency0 … 120 %
Air Ratio (Lambda)1 … 9.99
Excess Air0 … 99.9
CO/CO2 ratio0 … 10

O2 - Oxygen (Long Life)

Measuring range0 … 21%
Resolution0.1 %
Accuracy± 0.2 Vol-% abs.

CO - Carbon monoxide (H2 Compensated)

Measuring range0 … 10,000 / 20,000 ppm
Resolution1 ppm
Accuracy± 10 ppm o 5 % of reading < 4,000 ppm / 10 % of reading > 4,000 ppm

NO - Nitric oxide

Measuring range0 … 1,000 / 5,000 ppm
Resolution1 ppm
Accuracy± 5 ppm o 5 % of reading < 1,000 ppm / 10 % of reading > 1,000 ppm
Stack / Flue gas temperatureMeasuring Range: -40 … 2192 °F Resolution: 0.1 °F Accuracy: ± 4 °F … < 200°C / 1 % of reading > 200°C
Primary air / Ambient air temperatureMeasuring Range: 32 … 212 °F Resolution: 0.1 °F Accuracy: ± 4 °F
Differential temperatureMeasuring Range: -40 … 2192 °F Resolution: 0.1 °F Accuracy: ± 4 °F … or 0.5%
Stack draftMeasuring Range: +/- 80 inH2O Resolution: 1 inH2O Accuracy: ± 0.01 inH2O
Differential pressureMeasuring Range: +/- 80 inH2O Resolution: 1 inH2O Accuracy: ± 0.01 inH2O