- Technical details
- Description
- You May Also Like
- Your Recently Viewed Items
- Reviews
Technical details
Insulation resistance measurement:Measuring range according to IEC 61557-2VN= 15 000 V: 50 kΩ ... 40.0 TΩ
Ranges of measured resistance depending on the test voltage:
Capacitance measurement:
Measurement with DC and increasing voltage (SV) for VISO | Range | Resolution | Accuracy |
---|---|---|---|
5 kV | 0...999 kΩ | 1 kΩ | ±(3% m.v. + 10 digits) |
5 kV | 1.00..9.99 MΩ | 0.01 MΩ | ±(3% m.v. + 10 digits) |
5 kV | 10.0...99.9 MΩ | 0.1 MΩ | ±(3% m.v. + 10 digits) |
5 kV | 100..999 MΩ | 1 MΩ | ±(3% m.v. + 10 digits) |
5 kV | 1.00...9.99 GΩ | 0.01 GΩ | ±(3% m.v. + 10 digits) |
5 kV | 10.0...99.9 GΩ | 0.1 GΩ | ±(3% m.v. + 10 digits) |
5 kV | 100...999 GΩ | 1 GΩ | ±(3.5% m.v. + 10 digits) |
5 kV | 1.00...9.99 TΩ | 0.01 TΩ | ±(7.5% m.v. + 10 digits) |
≥10 kV | 10.0...20.0 TΩ | 0.1 TΩ | ±(12.5% m.v. + 10 digits) |
≥15 kV | 10.0...40.0 TΩ | 0.1 TΩ | ±(12.5% m.v. + 10 digits) |
Voltage VISO | Measuring range |
---|---|
50 V | 200 GΩ |
100 V | 400 GΩ |
250 V | 1.00 TΩ |
500 V | 2.00 TΩ |
1000 V | 4.00 TΩ |
2500 V | 10.00 TΩ |
5000 V | 20.0 TΩ |
10 000 V | 40.0 TΩ |
15 000 V | 40.0 TΩ |
Range | Resolution | Accuracy |
---|---|---|
0...999 nF | 1 nF | ±(5% m.v. + 5 digits) |
1,00...49,99 uF | 0,01 uF | ±(5% m.v. + 5 digits) |
- Displaying measured capacity after RISOmeasurement
- For measurement voltages below 100 V the measurement error is not specified
Reviews
Leave a review of this product!
If you've used this product, please leave a review to tell us and other customers what you thought about it. Get paid up to $16 for submitting one of the first text, photo, or video reviews for this item. Please contact us for more details.
Write a text review
Earn $2.00Post a product photo
Earn $4.00Post a product video
Earn $10.00