MRU Instruments 420082-03 AMPRO Plus Combustion / Emission Analyzer Kit, O2/CO/NO

  • Part #:420082-03
  • Model #:AMPRO plus
  • Condition:Brand New
51 Review Write a Review
    • $3,860.00 / Each
  • Free Shipping logoFree Shipping
See product details
Technical details
Max suction range gas pump / Typical gas flow150 hPa / 50 l/h
Internal memory32,000 data sets
Data transmission / InterfacesUSB, Bluetooth, WLAN / USB, Bluetooth, WLAN, IRDA, SD CARD
Display4" color display
Operation temperature / Storage temperature+5°C ... +45 °C / 41 °F ... 113 °F / -20°C ... +50°C / -4°F ... 122°F
Ambient conditions95% Rel. Humidity, non-condensing
Internal Battery Pack / Grid power supplyLi-Ion, 20h operation time / 100 – 240 V / 5V DC / 1200 mA
Protection classIP30
Dimensions / Weight4.3" x 8.8" x 2.04" (244x113x54 mm) / 1.65 lbs. (750g)
TÜV approvalTBD
Calculated valuesRange
Carbon dioxide0 ... CO2 max.
Heat losses qA0 ... 99.9 %
Efficiency0 ... 120 %
Excess Air0 ... 99.9
Air Ratio (Lambda)1 ... 9.99
CO/CO2 ratio0 ... 10
Other measured componentsMeasuring RangeResolutionAccuracy
Stack / Flue gas temperature32 ... 1,472°F (0 ... 800°C) with stainless steel
32 ... 2,012°F (0 ... 1100°C) with Inconel
1 °F± 4°F ... < 392°F / 1 % reading > 392°F
± 4°F ... < 392°F / 1 % reading > 392°F
Primary air / Ambient air temperature32 ... 212°F (0 ... 100 °C)1°F± 2°F
Differential temperatureUp to 2,012°F (0 ... 1100°C)1°F± 4°F ... < 392°F / 1 % reading > 392°F
Stack draft+/- 40 inH2O (100hPa)1 Pa± 0.02 inH2O or 1% reading
Differential pressure+/- 80 inH2O (200hPa)1 Pa± 0.02 inH2O or 1% reading
Gas flow velocity measurement3 ... 100 m/s (using Pitot tube)0.1 m/s